More Details

Contec SP950 Precise Infusion Syringe Pump

Contec SP950 Precise Infusion Syringe Pump : Product Description

The SP950 is a medical syringe pump adopting a microcomputer system and microelectronic technology—offering high intelligence and safety. The LCD displays parameters in real time and supports multi-function alarms (visual, auditory, and text prompts). It is easy to operate, stable, highly accurate, and automatically detects syringe specifications. The SP950 is widely used for IV injections across ICU/CCU, neonatology, obstetrics, veterinary and gynecology, pediatrics, internal medicine, surgery, operating rooms, and emergency departments.

Contec SP950 Precise Infusion Syringe Pump Features

  • User-settable infusion rate and total volume
  • Alarms for occlusion, syringe fall-off, communication abnormal, low battery, etc.
  • Adjustable screen brightness, alarm volume, and occlusion pressure range
  • Quick-feed function for rapid medicine supply
  • Automatic KVO mode after infusion completion
  • Prompt signal when the device remains in power-on state without operation
  • Built-in rechargeable battery with auto-switch power and auto-charge
  • Large screen with visual/auditory/text prompts for easy operation
  • Supports disposable syringes (CNS); accuracy ensured by parameter adjustment
Contec SP950 : Performance & Specifications

Performance

  • Supported syringe sizes (CNS): 10 ml / 20 ml / 30 ml / 50 ml
ItemSpecification
Infusion Rate (10 ml) 0.1–100 ml/h
Infusion Rate (20 ml) 0.1–200 ml/h
Infusion Rate (30 ml) 0.1–300 ml/h
Infusion Rate (50 ml) 0.1–500 ml/h
Quick-Feed Rate (10 / 20 / 30 / 50 ml) 100 / 200 / 300 / 500 ml/h
Accumulated Volume Display 0–999.9 ml
Accuracy ±3%
Occlusion Alarm Pressure (adjustable) Low: 40 kPa ± 16 kPa
Middle: 70 kPa ± 30 kPa
High: 100 kPa ± 35 kPa
KVO Rate 0.1–5 ml/h (adjustable)
Safety Classification Class II equipment, internally powered type BF
Alarm Functions ① Non-operated state  ② Near empty  ③ Empty (enter KVO)  ④ Syringe fall-off
⑤ Occlusion  ⑥ Pressure failure  ⑦ Motor failure  ⑧ Low battery  ⑨ Communication abnormal
Max Power Consumption 20–30 VA (continuous operation ≥ 2 hours at 5 ml/h after full charge)
Power Supply AC 100–240 V, 50/60 Hz
Battery Rechargeable Li-ion, 7.4 V, 3500 mAh
Working Environment Temp: +5–40 °C; Pressure: 860–1060 hPa; RH: 20–90% (no strong vibration/corrosive gas)
Storage Environment Temp: −20–+55 °C; RH ≤ 90%, no condensation; no corrosive gas; well-ventilated

Accessories (Box Contents)

  • Syringe Pump
  • Bracket
  • Power cord
  • User manual
  • Two fuses

Physical Characteristics

  • Dimensions (L × W × H): 310 × 140 × 135 mm
  • Weight: 1.7 kg