Bionet CardioCare 2000 12-Channel ECG Machine : Product Description

The CardioCare 2000 is an economically priced interpretive 12-channel ECG/EKG machine that ensures reliable performance without compromise. It is suitable for use in private practices, emergency rooms, and hospitals, designed for maximum ease of use and convenience.

Key Features

  • Simultaneous 12-channel resting ECG
  • ECG interpretation program
  • A4 full-size paper printing
  • Low maintenance cost
  • ECG filing system BMS-Plus
  • Upgradeable software

Specifications

ECG LeadsSimultaneous 12 leads resting ECG
Recording Channel3CH+1RHY, 6CH+1RHY, 12CH, 1CH long time (1 min)
Gain5, 10, 20, Auto (I~aVF: 10, V1~V6: 5) mm/mV
Printing Speed12.5, 25, 50 mm/sec
Sampling RateAnalysis: 500 samples/sec, Digital: 8000 samples/sec
FiltersAC (50/60Hz, -20dB), Muscle (25–35Hz, -3dB), Baseline drift (0.05–0.2Hz), Low pass (40–150Hz)
Display2 × 16 Char LCD (Gain, Speed, Filter, HR, Printing form, Rhythm)
User InterfaceShort Key
Printer ResolutionThermal print head, Roll paper, A4 Z-fold
Patient DataID, Name, Age, Gender, Height, Weight
Basic MeasurementHR (30–300 bpm), PR/RR int., QRS dur., QT/QTc int., P–R–T axis, SV1/RV5, R+S amp.
Noise & SafetyNoise ≤20uV, Input ≥50MΩ, Voltage ±5mV, CMRR >105dB, Defib protected
Line Power100–240V AC, 50/60Hz, Output 15VDC
BatteryReplaceable, Rechargeable Li-ion 11.1V, 2600mAh
Battery Capacity6 hours use / 200 ECG pages, recharge in 3 hours
CommunicationLAN
Safety ConformityClass I, Type CF defibrillation-proof
Environmental OperationTemp 10–40°C, Humidity 30–85%, Pressure 70–106kPa
Storage/ShippingTemp -20–60°C, Humidity 10–95%, Pressure 50–106kPa
Dimensions300(W) × 290(H) × 95(D) mm, 3.0 kg

Box Contents

  • ECG Cable ×1
  • Bulb Electrode ×1
  • Clamp Electrode ×1
  • Gel Bottle ×1
  • Power Cable ×1
  • Earthing Cable ×1
  • User Manual ×1

Additional Information

Warranty: 1 year