Nidhi Vision India Private Limited
Search

Cart

More Details

Lowenstein PrismaLab Bipap Features:

  • Modes: CPAP, APAP, AcSV, S, autoS, S/T, autoS/T, T, aPCV

  • Continued therapy and soft pressure transitions at change of mode during the night

  • Continuous check of mask fit, automatic mask recognition and mask test

  • High resolution therapy data for up to 14 days and detailed statistics for 366 days

  • Therapy software prismaTS for data analysis and remote control

  • With SD card and USB port

  • Telemedicine connection via modem for prisma CLOUD

  • Eight clearly identified analog signals for PSG feed

  • Intelligient humidification without extra sensors provided by smartAQUAcontrol (with use of prisma AQUA)

  • Comfort functions such as auto START-STOP and soft START

  • Heated breathing tube prisma HYBERNITE

  • Easy-to-use accessories

  • Lowenstein PrismaLab Bipap - Technical Specifications

Modes: CPAP, APAP, AcSV, S, autoS, S/T, autoS/T, T, aPCV
Mean Sound Pressure Level (ISO 80601-2-70): About 26.5 dB(A) at 10 hPa
CPAP/APAP Operating Pressure Range: 4 to 20 hPa
BILevel Operating Pressure Range: 4 to 30 hPa
AcSV Operating Pressure Range: 4 to 30 hPa
Relative Inspiration Time (Ti/Tset): 25% to 67%
Trigger: Auto or can be set at 3 levels
Pressure Increase Speed: Adjustable at three levels
Back-Up Frequency: Auto, 0 bpm to 35 bpm
Pressure Accuracy:
<20 hPa ± 0.6 hPa
≥20 hPa ± 0.8 hPa
Highest Flow (as per ISO 80601-2-70, Bilevel & AcSV modes, without humidifier):
4.0 hPa: 235 l/min
10.5 hPa: 225 l/min
17.0 hPa: 215 l/min
23.5 hPa: 200 l/min
25.0 hPa: 195 l/min
30.0 hPa: 190 l/min
Stability of Dynamic Pressure (Short-Term Accuracy) at 10 breaths/min (ISO 17510-1:2007, with 22-mm tube):
7.0 hPa: Δp ≤ 0.24 hPa
10.0 hPa: Δp ≤ 0.28 hPa
13.5 hPa: Δp ≤ 0.30 hPa
20.0 hPa: Δp ≤ 0.40 hPa
Dimensions (W x H x D): 170 x 135 x 180 mm
Weight: About 1.4 kg
System Interface: 12 V DC, max. 10 VA
Alarms:Leaks Disconnection (tube) 
Apnea Low minute volume Low tidal volume